Maintenance, Clean the air filter – MTS DCPD Measurement System User Manual

Page 40

Advertising
background image

Maintenance

Maintenance of the DCPD consists of keeping the unit clean and occasionally cleaning the filter.

The unit can be wiped off with a clean, damp cloth and then wiped dry with a clean, lint-free cloth.

Clean the Air Filter

The filter is stainless steel mesh and intended to be cleaned rather than replaced. It is recommended to
check the filter every 6 months (or less in dirty environments) and clean the filter if necessary. Perform the
following procedure to clean the filter.

Air filter

Description

Item

Air in. The filter is behind the cover.

1

Air out.

2

1. Disconnect the power cord.

2. Remove the four screws securing the right fan cover.

3. Remove the filter and clean it. The filter can be cleaned by blowing off with compressed air or cleaned

with soapy water, rinsed, and allowed to dry completely.

4. Replace the cleaned, dry filter.

5. Replace the cover removed in step 2.

6. Connect the power cord.

40 | DCPD Measurement System

Maintenance

Advertising