Controller configuration for dcpd – MTS DCPD Measurement System User Manual

Page 29

Advertising
background image

If prompted to enter a password, use the appropriate password.

The Open Configuration window will open.

b) Open the test config. file that was configured in the previous steps.. For example, "

Landmark_Frame1_DCPD.cfg".

The Station Manager window will open. All Station Manager windows that were previously opened
will open.

c) Open Station Setup.
d) Under Auxiliary Inputs, choose "ACT pd".
e) In ACT pd window, go to Limits.
f) Under the Excitation Failure tab, set Mode to "Disabled".
g) Open Station Setup.
h) Under Auxiliary Inputs, choose "REF pd".
i) In REF pd window, go to Limits.
j) Under the Excitation Failure tab, set Mode to "Disabled".
k) Set the drive limits.

1. Under Channels, select the Channel Drive icon.

2. In the Drive:Current window, set the Full Scale Minimum to -22.500 and the Maximum to

22.500

3. Set the Display Units to A for Amps.

10. Click Save in the Station Manager to save the settings.

11. Verify the installation with a Foil Specimen test.

Controller Configuration for DCPD

This section explains the typical controller configuration settings for a DCPD Measurement System. For
detailed information about controller configuration, refer to the MTS Series 793 Utility Software manual
and/or the MTS FlexTest Models 40/60/100/200 Controller Hardware manual.

DCPD Measurement System | 29

Installation

Advertising