MTS Model 680-01 High-Temperature Grips User Manual

Page 28

Advertising
background image

Model 680.01 High-Temperature Grips

28

6. Using spanner wrenches, rotate the two halves of the spiral washers in

opposite directions to tighten them. The opening between the washers must
not exceed an arc of 30° from the closed position.

If more than a 30° arc is created, remove the tensile load from the specimen.
Remove the specimen and tighten the grips on the mounting studs, taking
care not to misalign the wedge openings. Repeat Step 4 through Step 6 of
this procedure.

7. Reduce tensile force to zero and remove the specimen.

8. Remove the system hydraulic pressure.

Connector Stud and Spiral Washers

Connector Stud

Decrease Thickness

Medium Thickness

Do not exceed.

Connector
Stud

Increase Thickness

Advertising