Operation, Determine the gripping pressure, Operation 33 – MTS Model 680-01 High-Temperature Grips User Manual

Page 33: Determine the gripping pressure 33

Advertising
background image

Model 680.01 High-Temperature Grips

33

Operation

This section describes how to use the grips.

Contents

Determine the Gripping Pressure 33

Install Threaded Specimens 35

Install Button-End Specimens 37

Adjust the Water Flow 38

Remove Specimens from the Grips 39

Determine the Gripping Pressure

The amount of hydraulic pressure to be applied to the grips must be sufficient to
prevent backlash between the specimen ends and the specimen adapters, and also
between the mating threads of the specimen adapters and the grip extensions,
during all parts of the loading cycle.

To also ensure that the grips will not be loaded beyond their force rating,
however, determining hydraulic pressure must start with temperature
considerations, rather than the forces to be applied to the specimen.

Note

If the gripped ends of the specimen become deformed, reduce the
hydraulic gripping pressure. Re-compute the amount of pressure
required to grip the specimen.

Determining the required hydraulic pressure is a two step process:

1. Determine that the maximum force (whether tensile or compressive) to be

applied to the specimen is within the grip rating, considering the
temperatures that the specimen will experience.

2. If anticipated forces are appropriate, use the “Required Pressure for the

Applied Force” graph to find the required gripping pressure.

Force versus

temperature

The following figure shows that the grip force rating for specimen temperatures
up to 700°C is 67 kN (1292°F at 15,000 lb) is stable. The figure also shows that
the force rating is reduced proportionately for specimen temperatures between
700°C and 1000°C (1292°F and 1832°F), with a maximum force rating of 8.9 kN
at 1000°C (2000 lb at 1832°F).

Advertising