Install threaded specimens, Install threaded specimens 35 – MTS Model 680-01 High-Temperature Grips User Manual

Page 35

Advertising
background image

Model 680.01 High-Temperature Grips

35

Install Threaded Specimens

Perform the following procedure to mount threaded specimens in the grips.
Ensure that all grip parts are sufficiently cool before touching them.

Specimen mounting is a potentially hazardous procedure because hydraulic
pressure is applied to the system and actuator movement is required.

This can cause personal injury. It can also apply damaging forces to the
specimen, furnace, or to other fixtures.

Use extreme care while performing the following steps.

1. Determine the hydraulic pressure that must be applied to the grips in order

to adequately grip the specimen.

2. Ensure sure that the specimen adapters are clean and free of accumulated

oxide deposits.

3. Apply a liberal amount of anti-galling compound (supplied with the grips)

to the threads of both specimen adapters and to the threads of the specimen
to ensure easy removal after high-temperature operation.

4. Screw one end of the specimen into one of the specimen adapters until it

extends through the adapter.

Threaded Specimen Installation

5. Screw the specimen adapter fully into the grip extension of the upper grip.

6. While looking into the viewing port in the grip extension, screw the

specimen into the specimen adapter as required to center the end of the
specimen in the viewing port.

7. Screw the other specimen adapter fully into the extension of the lower grip.

WARNING

Upper Grip Extension

Viewing Port

Specimen Adapter

Threaded Specimen