Connecting the hydraulic lines – MTS Series 505 SilentFlo Hydraulic Power Unit Model 505-30 User Manual

Page 33

Advertising
background image

505.20/.30 HPU Connections

Model 505.20 - 505.30 SilentFlo™ HPU

Installation

33

The first time a pump motor is started after electrical service is connected to
the HPU, it is possible for the pump motor to rotate in the wrong direction.

A pump motor rotating in the wrong direction at high pressure for longer
than 10 seconds can cause severe damage to your HPU.

If the output pressure does not increase within 10 seconds after High is selected,
stop the unit.

2. Connect the grounding wire to the lug labeled PE GND (protective earth

ground).

Connecting the

hydraulic lines

The figure shows the hydraulic adapter ports to the HPU.

Note

Each hydraulic connection requires an O-ring face seal.

CAUTION

PE GND (Protective
Earth Ground) Lug

Power
Disconnect

Electrical Power Feed
Lugs (shown labeled for
counterclockwise phase
orientation)

Ensure lugs are secure.
They should be checked
periodically. Refer to the
Maintenance section.

HPU Power Requirement
and Power Feed Phase
Orientation Identifier

Return
-16 (1 in)

Drain
-8 (1/2 in(

Water In
1 in NPT

Pressure
-16 (1 in)

Drain
-6 (3/8 in)

Water Out
1 in NPT

Advertising
This manual is related to the following products: