MTS Insight Material Testing Systems User Manual

Page 44

Advertising
background image

MTS Insight™ Material Testing System

44

Controller Connections

Installation

Transformer Requirements for MTS Insight 150, 200 and 300 Machines

Important

Remember to verify local codes, voltages and frequencies before
ordering transformer.

Following are the minimum requirements for transformers for the MTS Insight
150, 200, and 300 load frames.

Isolation Type Transformer (Electrical isolation between input and output).

3-phase input and output.

Wye output with neutral connection at 400 V AC.

Equi-potential between each output phase and earth/ground.

Ambient operating temperature from 5–40° C.

Temperature rise 150 deg C maximum.

Insulation class 200 deg C minimum.

Electrostatic Shield.

UL Listed along with CE and CSA/CUL or equivalent depending country.

Additional

considerations

Floor standing.

NEMA 2 or better enclosure.

Harmonic Factor can be 0.

Maximum sound level 45 dB per ANSI Standard C89.2.

Typical power in

selected areas

United States - Voltage and Frequency

480 VAC @ 60 Hz

Canada - Voltage and Frequency

600 VAC @ 60 Hz

Europe - Voltage and Frequency

400 VAC @ 50 Hz