Appendix, Additional digital i/o information, Appendix 73 – MTS Insight - 1kN to 300kN User Manual

Page 73: Additional digital i/o information 73

Advertising
background image

Additional Digital I/O Information

Insight Material Testing System

Appendix

73

Appendix

Additional Digital I/O Information

The digital inputs have an MOCD223 optical isolator with 2.7 Kohm,
½ Watt series resistor. To reliable turn on they need 1 mA of current.
This means the minimum input high voltage is 4.0 Vdc. The
maximum input voltage is 28.0 Vdc. The device should be off for
input voltages less than 1.0 V.

The digital outputs are implemented by an AQV252G PhotoMOS relay with
a 0.75 Amp poly fuse in series. Although the device is rated at 60 V peak but
it is recommended that a maximum of 48 V be applied. If the load is highly
inductive, such as a relay coil, an appropriate snubber network should be
used near the coil terminals to prevent large flyback voltages from exceeding
the device ratings.

Insight Controller

User
Digital
Output

User
Digital
Input

From Logic
Controller

To Logic
Controller

Optical
Isolator

PhotoMOS
Relay

Advertising