Model 505.07/.11 specifications – MTS Landmark Testing Systems User Manual

Page 18

Advertising
background image

Landmark Test System Site Prep Guide

18

Hydraulic Power Unit Specifications

System Component Specifications

Model 505.07/.11 Specifications

General Specifications

Parameter

Specification

Environmental

Operating temperature

Humidity

Heat load

Noise

*

rating at 1 m

* Sound pressure level [db (A)] is expressed as a free field value. Readings may vary with

the acoustic environment.

For use in a controlled environment

5–40°C (41–104°F)

0–85% noncondensing

<630 kcal/hr (2500 BTU/hr)

60 dB (A) fully compensated

Hydraulic fluid

Typical operating
temperature range

Filtration

Reservoir capacity

Exxon Mobil DTE 25 or Shell Tellus

46 AW

43–49°C (110–120°F)

3 microns nominal

174 L (46 gal) maximum

Heat exchanger

Flow requirements

Water pressure

Maximum pressure

Cooling

Water hose

Water-cooled stainless steel plate

See “Water flow rating (input temperature)”

0.24–0.34 MPa (35–50 psi differential)

0.83 MPa (120 psi)

see the Model Specific Specifications table

19 mm (0.75 in)–inside diameter

Hydraulic connections

Pressure

Return

Drain

Each requires an O-ring face seal

-12

-12

-8 and -6 (two connections)

Water Connections

-12 (3/4 in NPT) for both In and Out

Electrical power

Line voltage

Control voltage

200–575 V AC, 3

∅ at 50/60 Hz

24 V DC

Dimensions

Height

Length

Width

1067 mm (42 in)

990 mm (39 in)

711 mm (28 in)

Advertising