Controller specifications, Environmental requirements—series 494 hardware – MTS Landmark Testing Systems User Manual
Page 22
Landmark Test System Site Prep Guide
22
Controller Specifications
System Component Specifications
Controller Specifications
Environmental Requirements—Series 494 Hardware
All Series 494 hardware components are intended for indoor use only. This
indoor environment must conform to the following environmental specifications.
Note
All Series 494 Controller must only be operated under the installation
and ambient conditions (such as, temperature, moisture, and EMC)
specified.
Note
To maintain EMC compliance, the controller must be installed in a
location that does not exceed the EN 61000-6-4 emission standard for
industrial environments.
Parameter
Specification
Temperature
5ºC–40ºC (41ºF–104ºF)
Humidity
5–85%, non-condensing
Altitude
3048 m (10,000 ft) maximum
Space
Requirements
For proper ventilation, allow 51 mm (2 in) clearance
on all sides of the chassis.
The rear of the chassis requires a minimum
clearance of 15.24 cm (6 in) for cable connections.