Specifications–model 494.04 chassis – MTS Landmark Testing Systems User Manual

Page 23

Advertising
background image

Controller Specifications

Landmark Test System Site Prep Guide

System Component Specifications

23

Specifications–Model 494.04 Chassis

All equipment related to the controller should be connected to the same fused
power circuit.

Note

Electrical connections must be made by qualified personnel and conform
to local codes and regulations. Local electrical codes supersede any
information found here.

Model 494.04 Chassis Specifications

Parameter

Specification

Input Voltage

100–240 V AC (single phase)
power factor corrected universal input

Input Frequency

50–60 Hz

Input Current

5 A at 100 V AC

3 A at 240 V AC

Facility Power Requirements

Provide a dedicated circuit for the
chassis, computer, and monitor.

Input Surge

<40 A

Insulation Over Voltage

Category I

Pollution Degree

2

Weight

Approximately 8.6 kg (19 lb)

Dimensions

Width: 43 cm (17 in)
Height: 14 cm (5.5 in)
Depth: 44.5 cm (17.5 in)

Advertising