Environmental requirements – MTS Acumen Electrodynamic User Manual

Page 29

Advertising
background image

Electrical connections must be made by qualified personnel and conform to local codes and regulations.
Local electrical codes supersede any information found here.

MTS Acumen 3

MTS Acumen 1

Electrical Requirements

200-240 VAC

100-120 VAC (200-240)

Voltage

50-60 Hz

50-60 Hz

Frequency

10 Amp

7 (4) Amp

Current

12

Single

Single

Phase

Environmental Requirements

All MTS Acumen load frames are intended for indoor use only. This indoor environment must conform to the
following environmental specifications.

Environmental Requirements

Specification

Parameter

5° to 40° C (41° to 104° F)

Operating Temperature

13

5 to 85% Non-condensing

Operating Humidity

-18° to 49° C (0° to 120° F)

Storage Temperature

90% Non-condensing

Maximum Storage Humidity

2000 m (6562 ft)

Maximum Altitude

2

Pollution Degree

IP20

IP Rating

II

Overvoltage Category

12

MTS Acumen 1 current rated at 100 VAC. MTS Acumen 3 current rated at 200 VAC.

13

Take care when using heat producing accessories, such as ovens. Emitted heat from such devices may
damage the MTS Acumen if not properly controlled.

MTS Acumen™ Electrodynamic Test System 29

System Specifications