Site location, Transporting – MTS Acumen Electrodynamic User Manual

Page 32

Advertising
background image

Site Location

Before installation of the MTS Acumen system, ensure the final location meets the following requirements.

The floor and table are able to support the weight of the frame and its computer components.

The environment complies with the specifications for your specific model:

Operating temperature

Storage temperature

Humidity

Atmosphere

The required power supply is available and is compatible with the electrical requirements for your specific
model.

Grounded electrical outlets are within the minimum distance requirement to the load frame location.

A phone line or cell phone access is located within the general testing area so that users can call the MTS
Service department directly from the testing area. It is also suggested that network drops, or digital phone
lines, be within the general testing area.

Ceiling clearance is adequate for the load frame, including the additional space necessary to lift and move
the frame by either a forklift or crane.

Location provides sufficient accessibility for routine maintenance: 1000 mm (3 ft).

Employees are adequately trained to operate the load frame and its computer system.

Transporting

Before transporting the MTS Acumen system, ensure the following requirements are met.

You have the correct number of packing boxes as stated on the packing list.

Insurance coverage is secured or verified.

You have the required equipment for the chosen transporting method.

You have the appropriate packaging to protect the load frame when moving or relocating it.

The pathway from the loading dock to the final site location has sufficient width and height to fit the frame
and forklift (or crane).

The floor area along the pathway to the final site location is able to support the weight of the frame and
forklift (or crane).

There are no loose accessories on the shipping skid or frame.

Your equipment operators have the appropriate licenses and have complied with your local safety standards
(for example, the appropriate training required by OSHA in the U.S.).

32 MTS Acumen™ Electrodynamic Test System

Site Requirements Checklist