Electrical Power Requirements

MTS Criterion

®

Series 40 Site Preparation Guide

System Specifications

21

Electrical Power Requirements

The following table provides the electrical power requirements of various load
frame types

.

Note

Electrical connections must be made by qualified personnel and conform to local
codes and regulations. Local electrical codes supersede any information found
here.

Electrical Power Requirements

M

ODEL

V

OLTAGE

F

REQUENCY

C

URRENT

M

AXIMUM

P

OWER

R

EQUIREMENT

C41
C42

Single phase
200-230 VAC

50/60 Hz

5 A

1000 W

C43 (10 kN)
C44 (10 kN)

Single phase
200 - 230 VAC

50/60 Hz

10 A

2000 W

C43 (30 kN, 50 kN)
C44 (30 kN)
C45 (50 kN)

Single phase
200 - 230 VAC

50/60 Hz

12 A

2400 W

C45 (100 kN)

Single phase
200 - 230 VAC

50/60 Hz

22 A

4400 W

C45 (300 kN)

Three phases
380-480 VAC

50/60 Hz

20 A

5200 W

C45 (600 kN)

Three phases
380-480 VAC

50/60 Hz

20 A

5200 W

Facility Power
Requirements:

Provide a dedicated circuit for the machine, computer, and monitor.

Insulation Over Voltage

Category II

Pollution Degree

2