Electrical Power Requirements
MTS Criterion
®
Series 40 Site Preparation Guide
System Specifications
25
C45 (300 kN,600 kN)
step up transformer
C45 (300 kN,600 kN)
step down transformer
Transformer Connection for C45(300 kN, 600 kN)
208, 60 Hz
3- Phase VAC
Wall Power
Transformer
100-249-050
MTS Attached
Power Cord
Supplied
Wall Switch
600, 60 Hz
3- Phase VAC
Wall Power
Transformer
100-249-048
MTS Attached
Power Cord
Supplied
Wall Switch
Transformer Connection for C45(300 kN, 600 kN)