MTS Landmark TWX User Manual

Page 108

Advertising
background image

4. Release the upper grip.

Use the controls on the load frame.

5. Retract the actuator.

a) In the Station Manager application main window, click the Manual Command button to open the

Manual Command window.

b) Select the following parameters:

Channel: Ch1

Control Mode: Displacement

Enable Manual Command: Checked

c) Using the slider control, slowly retract the actuator until you have enough clearance to remove the

specimen.

6. Turn off the HSM.

108 | MTS Landmark Test System Operation

Running the Example Tension Test