In the Station Manager application main window, click the Reset button to clear the interlock.

4. Re-power the HPU and HSM.

In the Station Manager application main window, click the Power Low button, and then the Power High
button for the HPU and HSM.

5. Reset Force control to zero.

In the Station Manager application main window, click the Manual Control button. In the Manual Controls
window, reset Manual Cmd to 0.

6. Reset the meters.

Reset meter history by selecting Station Manager > Display > Meters and clicking the Reset button.

7. Reset the Lower (Compression) Limit.

112 | MTS Landmark Test System Operation

Running the Example Tension Test