Excessive I Gain

a) Navigate to the Tuning: Axial Force window found by selecting Station Manager > Display > Station

Setup > Channels > Ch1 (or other depending on your configuration) > Force, clicking the Tuning
Fork button, and clicking the Adjustments tab.

b) Compare your graph to the examples of I gain found in the following displays. If necessary, adjust

the I gain for the Force channel.

Note: Do not adjust all the way to zero or you could trip the force limit.

11. Turn off the Function Generator.

MTS Landmark Test System Operation | 93