Ttl i/o operational trigger default configuration, Ttl i/o directional control input configuration – New Era NE-4000 Multi-Phaser Programmable Double Syringe User Manual

Page 20

Advertising
background image

New Era Pump Systems Inc.

www.SyringePump.com

Model NE-4000 Multi-Phaser™

Publication #1200-01

15

09/05/14


7.4

TTL I/O Operational Trigger Default Configuration

aa

Configures the default mode of the TTL I/O ‘Operational Trigger’ (pin 2) that will control the

Pumping Program’s operation. This default setting can be overridden within a Pumping Program. (See sec. 10.1,
TTL I/O Operational Controls

). The 2 letter configuration Setting parameter to the right of the colon (:) is defined

in column ‘Setting’ as follows:

Setting

RS-232 Command Setting

FUN TRG <n>

<n> =>

Name

Function

Ft

0

Foot Switch

Falling edge starts or stops the Pumping Program

FH

1

Foot Switch Hold

Falling edge starts the Pumping Program
Rising edge stops the Pumping Program

F2

2

Foot Switch Reversed

Rising edge starts or stops the Pumping Program

LE

3

Level Control

Falling edge stops the Pumping Program
Rising edge starts the Pumping Program

St

4

Start Only

Falling edge starts the Pumping Program

t2

5

Start Only Reversed

Rising edge starts the Pumping Program

SP

6

Stop Only

Falling edge stops the Pumping Program

P2

7

Stop Only Reversed

Rising edge stops the Pumping Program

rL

8

Start on Low Level

Start Pumping Program on a low level

rH

9

Start on High Level

Start Pumping Program on a high level

SL

10

Stop on Low Level

Stop Pumping Program on a low level

SH

11

Stop on High Level

Stop Pumping Program on a high level

oF

12

Trigger off

Disable trigger

Et

13

Trigger Event

Pumping Program Function Only:. Trigger an
event instead of stopping the pump

bt

14

Button Trigger Event

Pumping Program Function Only: Pressing ‘Stop’
key triggers an event instead of stopping the pump

7.5

TTL I/O Directional Control Input Configuration

aa

Configures how the TTL input ‘Pumping Direction’ (pin 3) will control the pumping direction. (See

sec. 10.1, TTL I/O Operational Controls). The 2 letter configuration parameter to the right of the colon (:) is
defined as follows:

Setting Name

Function

rE

Reciprocating Pumps

Falling edge: Infuse; Rising edge: Withdraw

dU

Dual Pump

Falling edge: Withdraw; Rising edge: Infuse

The setting names are relevant to a 2 pump system, whereby the ‘Directional Control Input’ TTL pin is attached to
the second pump’s ‘Pumping Direction Output’ TTL pin.

In addition, “Directional Control Input” will affect how the “Sticky Direction” function operates. See “ Sticky
Direction”, sec: 5.8.1