RS-232 Pump Network Connector Wiring

RS-232

To Network To Computer


PC Com Port Connectors

25 Pin

9 Pin

3 - Receive

2 - Receive

Connect to pump TXD

2 - Transmit

3 - Transmit

Connect to pump RCV

9 - Ground

5 - Ground

Connect to pump GND

11.3

Accessories

11.3.1

ANA-BOX™

Part #: ADPT-ANABOX, Analog voltage control interface.

Allows the pump to be controlled by a variable voltage source. Allows the Creation of a closed loop system
allowing the NE-4000 to be controlled by an external sensor, such as a pressure sensor, or other variable voltage
source. Start or stop the pump at a specific voltage level. Set the pumping rate to be proportional to the voltage
input.

11.3.2

Syringe Heater

Part#: SYRINGE-HEATER, Visit

www.SyringeHeater.com

for details.

Flexible heating pad that wraps around the syringe. Thermo-Kinetic Heat Clamping digital controller will heat a
syringe to a set temperature up to 100 C.

11.3.3

RS-232 Network Cables

RS-232 Network Primary Cable
7 foot cable, part #:

CBL-PC-PUMP-7

25 foot cable, part #:

CBL-PC-PUMP-25

Cable to connect a pump, or the first pump in a pump network, to a standard personal computer’s serial port with a
DB-9 or DB-25 connector. Included with this cable is a 9 pin to 25 pin converter.

RS-232 Network Secondary Cable
7 foot cable, part #:

CBL-NET-7

25 foot cable, part #:

CBL-NET-25

Cable to connect additional pumps, after the first pump, to the pump network.

USB to RS-232 Converter Cable
USB to RS-232 cable, software drivers on CD, part#: CBL-USB232
Attached to the RS-232 Network Primary Cable, allows communication through a PC’s USB port.

11.3.4

Automation Cable: Special Communications Modes Control Cable

Part #: CBL-DUAL-3

Using two NE-4000 syringe pumps, this cable either creates an automated, continuous operation pumping system,
whereby one pump is refilling while the other is dispensing, or creates a dual pumping system, with both pumps
operating in the same direction. Plus other special communications modes.