The Trigger Event Trap is a one-time function. After triggering an Event Trap, the Trigger Event Trap is cleared.
The Trigger Event function will not affect the current trigger configuration.

8.3.19

'OUt.n': Set TTL Output Pin

n

Set ‘Program Output’ TTL I/O output (pin 5) to level ‘n’. If ‘n’ = 0, the output pin will be set low. If

‘n’ = 1, the output pin will be set high.

8.3.20

'bEEP': Beep

Sounds a short beep.

8.4

Pumping Program Examples

8.4.1

Example 1: 2 Step Rate

ml Dispensed

0 5

30

0

2.5

500

m

l/

h

r

R

a

te

Infuse 5.0 mL at 500 mL/hr, and then infuse 25.0 mL at 2.5 mL/hr. Then stop the pump.

Phase

Function

Rate

Volume

Direction

1

RATE

500 mL/hr

5.0 mL

Infuse

Phase

Function

Rate

Volume

Direction

2

RATE

2.5 mL/hr

25.0 mL

Infuse

Phase

Function

3

STOP

8.4.2

Example 2: Repeated Dispenses with Suck Back

m

l/

h

r

R

a

te

Time

0

0

750

5:00

:30

Dispense 2.0 mL with a 5 minute pause between dispenses. In addition, after each dispense, a volume of 0.25 mL
is sucked back to prevent dripping. Also, 30 seconds before the end of the pause interval, a beep is sounded to alert
the operator to prepare for the next dispense.

Starting with the second dispense, 0.25 is added to the volume dispensed to compensate for the sucked back
volume of the previous dispense. By changing the last Phase to a [LP:nn] function, the total number of dispenses
can be set.

When entering a function with associated data, such as with the ‘Pause’ in Phase 5, or the ‘Loop’ in Phase 6, the
function is entered in 2 steps. First select the function and store it. Then enter the associated data.

Phase

Function

Rate

Volume

Direction

1

RATE

750 mL/hr

2.0 mL

Infuse

Phase

Function

Rate

Volume

Direction

2

RATE

750 mL/hr

0.25 mL

Withdraw

Phase

Function

3

LP:ST

Phase

Function

4

LP:ST