New Era NE-4000 Multi-Phaser Programmable Double Syringe User Manual

Page 42

Advertising
background image

New Era Pump Systems Inc.

www.SyringePump.com

Model NE-4000 Multi-Phaser™

Publication #1200-01

37

09/05/14


VOLUME DISPENSED

DIS

Queries volume dispensed only. Set not applicable.

Response:

I <float> W <float> <volume units>

Where: “I <float>“ refers to the infusion volume dispensed, and “W <float>” refers to the withdrawn
volume.

CLEAR VOLUME DISPENSED

CLD { INF | WDR }

Sets the Infused or withdrawn volume dispensed to 0. Command is only valid while the Pumping
Program is not operating. Query is not applicable.

INF

= Infusion volume

WDR

= Withdrawn volume

Query is not applicable.

9.4.3

Configuration and Setup Commands

New settings for any of the following commands will be stored in the non-volatile memory.

LOW MOTOR NOISE

LN [ <on-off> ]

Set/query low motor noise mode. Increases motor stepping resolution.

ALARM SETUP

AL [ <on-off> ]

Set/query alarm setup mode. Set alarm enables or disables alarm buzzer mode.

POWER FAIL SETUP

PF [ <on-off> ]

Set/query Power Failure mode. Set Power Failure enables or disables Power Failure mode.

TTL I/O OPERATIONAL TRIGGER DEFAULT SETUP

TRG [ <trigger setup> ]

Set/Query TTL I/O Operational Trigger (Pin 2) default input configuration.

TTL I/O Operational Trigger is set to <trigger setup>.

<trigger setup> =>

FT

= Foot switch trigger (falling edge start/stop)

FH

= Foot switch hold (falling edge start, rising edge start)

F2

= Foot switch reverse (rising edge start/stop)

LE

= Level trigger (rising edge start, falling edge stop)

ST

= Start only trigger (falling edge start)

T2

= Start only trigger reversed (rising edge start)

SP

= Stop only trigger (falling edge stop)

P2

= Stop only trigger reversed (rising edge stop)

RL

= Start on low level

RH

= Start on high level

SL

= Stop on low level

SH

= Stop on high level

OF

= Trigger off (disabled)