New Era NE-4000 Multi-Phaser Programmable Double Syringe User Manual

Page 30

Advertising
background image

New Era Pump Systems Inc.

www.SyringePump.com

Model NE-4000 Multi-Phaser™

Publication #1200-01

25

09/05/14


Phase

Function

Rate

Volume

Direction

10

INCR

1.0

0.1 mL

Infuse

Phase

Function

11

LP:50

Phase

Function

12

JP:02

8.4.4

Example 4: Complex Dispenses with External Synchronization

0

0

750

W

it

h

d

ra

w

(

m

l/

h

r)

I

n

fu

s

e

Start
Trigger

Start
Trigger

:60

:60

:60

Start
Trigger

Start over

Syringe

Refill

A more complex dispensing example, this Program contains different pumping requirements, including dispenses
with multiple pumping rates. The first set of 3 dispenses drops down to a lower pumping rate during the dispense.
When each dispense is completed, the buzzer beeps to alert the operator, then the pump waits for a start trigger
before starting the next dispense.

The next set of 3 dispenses have a fixed time interval of 60 seconds between dispenses. After the last set of
dispenses, the syringe is refilled by the amount infused, 17.25 mL. Then the buzzer beeps, to alert the operator to
the start of the first set of dispenses. The process is then repeated.

Phase

Function

Rate

Volume

Direction

1

RATE

750.0 mL/hr

0.5 mL

Infuse

Phase

Function

Rate

Volume

Direction

2

RATE

300.0 mL/hr

1.5 mL

Infuse

Phase

Function

3

BEEP

Phase

Function

4

PS:00

Phase

Function

5

LP:02

Phase

Function

Rate

Volume

Direction

6

RATE

750.0 mL/hr

0.5 mL

Infuse

Phase

Function

Rate

Volume

Direction

7

RATE

300.0 mL/hr

1.5 mL

Infuse

Phase

Function

8

BEEP

Phase

Function

9

LP:ST

Phase

Function

10

PS:60

Phase

Function

Rate

Volume

Direction

11

RATE

500.0 mL/hr

3.75 mL

Infuse