RaySafe i2 Dose Manager User Manual

Page 13

Advertising
background image

13

RaySafe i2 dose manager user manual – Get to know dose manager

FUNCTION

DEsCRIpTION

displayed name

the name that is displayed when the dosimeter appears in a real time
display. this name is also used to identify a dosimeter in dose viewer
and dose manager. this text field is limited to maximum 16 characters
due to space limitations in the real time display.

dosimeter position

Reflects where on the person the dosimeter is positioned and can be
set to one of Head, torso, Hand, belly, leg or other.

Clinical role

Reflects the role of the person using the dosimeter and can be set to
one of doctor, nurse, technician or other.

displayed symbol

the color which is displayed when the dosimeter appears in a real time
display.

Annual dose limit

this value is used to show the annual dose value in relation to the
annual dose limit in percent. If you do not want to show this relation,
set this value to 0.

NOTE

The annual dose limit for the dosimeter is dependent of the shielding factor of the lead apron,
as well as other radiation protection used. The more protection the higher the limit can be for
the same effective dose to the user. The annual dose limit may or may not be chosen to reflect
the legal dose limit where it is used. It could reflect the legal limit or for example a lower
target for the clinical user case. These and other considerations need to be counted for when
choosing the annual dose limit.