RaySafe i2 dose manager user manual – Get to know dose manager

FUNCTION

DEsCRIpTION

Power mode

• on – Communication with real time displays will take place and

registration of dose data will occur

• off – Power saving mode. no communication with real time

displays will take place and no registration of dose data will occur.

dose history

the accumulated dose history, the dose rate history and dose Sessions
1 and 2 will be reset to 0. the following happens when the dose history
is reset:

• In dose viewer and real time displays it will appear as if the

dosimeter has started measuring from 0 again. the deleted dose
history cannot be retrieved.

• In dose manager, the dose history that was available before reset

will be archived and possible to retrieve by clicking + next to the
dosimeter in the All dosimeters window. dose manager will start
a new session for the dosimeter and it will appear as if the dose
history will start measuring from 0 in the new session.

technical information

battery status indicates the dosimeter’s battery status:

• Green and yellow indicates normal use.

• Red indicates that the dosimeter need to be replaced in 4-6 months

at normal use.

• A crossed battery indicates that there is no battery left. the

dosimeter does not measure radiation and will not communicate
with real time displays.

firmware version refers to the firmware that is currently installed in the
dosimeter.