6 safety functions and fault display, Safety functions and fault display, Sample receptacle – Retsch BB 50 User Manual
Page 27: Fault messages
![background image](/manuals/571037/27/background.png)
Safety functions and fault display
27
The red LED above the STOP button lights up and the green LED above the
START button goes out. The machine stops.
Pos : 9. 27 /0020 Übersc hrift en/1.1 Ü bersc hriften/ 1. 1 Ü berschrif ten BDA/11 Probenbehälter @ 5\ mod_1345454254753_9. doc x @ 34220 @ 2 @ 1
5.12
43B
Sample receptacle
Pos : 9. 28 /0005 RET SCH/ 0005 R ETSCH Bedienungsanlei tungen Kapit elsamml ung en/BB50/ 0007 Bedi enung/ 0016 BB50 M odul Probenbehäl ter @ 5\ mod_1346050971062_9.doc x @ 34683 @ @ 1
Fig. 21: Sample receptacle
Pos : 11.1 /0020 Übersc hrift en/1. Übersc hrift en/1 Sic herheitsf unktionen und F ehl eranz eige @ 0\ mod_1226908695377_9. doc x @ 3594 @ 1 @ 1
6 Safety functions and fault display
Pos : 11.2 /0020 Übersc hrift en/1.1 Ü bersc hriften/ 1. 1 Ü berschrif ten BDA/11 F ehler mel dungen @ 0\ mod_1222344568006_9.doc x @ 2435 @ 2 @ 1
6.1
44B
Fault messages
Pos : 11.3 /0005 RET SCH/ 0005 R ETSCH Bedienungsanlei tungen Kapit elsamml ung en/BB50/ 0009 Sicherheits funktionen und Fehler anz eige/0008 BB50 M odul F ehler mel dungen @ 5\ mod_1346844648119_9.doc x @ 34851 @ @ 1
E
rro
r
c
ode
DESCRIPTION
D
is
pl
ay
gap w
idt
h
D
is
pl
ay
s
peed
E 10 DRIVE OVERLOADED
E 10
Switch machine off and back on. Where
necessary wait 10 minutes
E 22 ERROR KEYPAD
E 22 Switch machine off and back on.
E 26
ERROR FREQUENCY
CONVERTER
E 26
Frequency converter is faulty. Service is
required
E 50 ERROR SAFETY CIRCUIT
E 50
Safety circuit is faulty. Service is
required
E 80 ERROR INTERFACE
E80
Communication fault in controller
Service is required
H 41 CLOSE GRINDING CHAMBER
H41 Close hopper/drawer and press STOP
H 43 WEAR LIMIT REACHED
H 43 Replace breaking jaws
H 44 PLEASE CALIBRATE
PLS
CAL Calibrate gap width
"Gap width" und "Speed" displays flash
88.8
888
Close hopper and drawer;
restart the machine