9 fault messages, G1 g4 g5 g6, Fault messages – Retsch SM 200 User Manual

Page 37

Advertising
background image

Fault messages

37

Pos : 15.2 /00005 Ü berschrif ten/ 1. Ü berschrif ten/ 1 F ehler mel dungen @ 2\ mod_1259056589728_9.doc @ 17580 @ @ 1

9 Fault messages

Pos : 15.3 /00010 Bedi enungsanleit ung en Kapitels ammlungen/ SM200/ 0035 SM 200 Sic her hei tsf unktionen und Fehler anzeige3505 SM200 M odul F ehl er mel dungen ( 2013- 01-29 09: 01: 30) @ 3\ mod_1280233554638_9.doc @ 22214 @ @ 1

Flashing sequence

Meaning

Instruction

G4 and G5 flash

alternately

Safety circuit is not
closed

Close the door.

Pull the locking bracket forwards.

Allow the locking bolt to engage.

G5 flashes

Motor block

Remove the blocking pieces of sample material
from the grinding chamber.

Press button G6

Press button G1.

G4 + G5 flash

simultaneously

Faulty relay contact

Service necessary!

G4 flashes

Revolving field
detection

Switch the two phases (L1 with L2) on the power
cable

Fig. 30:Flashing sequence during error message