Retsch AS 200 tap 30.025.0001 User Manual

Page 16

Advertising
background image

08.01.2008

Retsch GmbH

16

Doc.Nr. D 98.025.9999

Starting – Interrupting - Stopping

15 16

Starting:

Press START / button 15 and M simultaneously

Green LED over button 15 lights up

Interrupting (pause function):

Press STOP / button 16 once
Red LED over button 16 lights up
Values remain visible
Press START / button 15 and M simultaneously

Sieving process is continued

Stopping (stand-by function):

Press STOP / button 16 twice
Red LED over button 16 lights up
The complete display is extinguished
Activate displays = press button 15 once
New values can now be eintered
Press START / button 15 a second time

Function as with Starting is carried out

Inserting the sieve stack

The AS 200 tap is suitable for analytical sieves of
200 mm and 203 mm (18“) outside diameter.

In order to loosen the clamping screws easily you

can change the position of the clamping levers
KH independently of the clamping screws. To do
this pull the clamping levers forwards and turn
them both to the required position.

·Then unscrew the two clamping screws C and

push the carriage downwards. Tighten both
clamping screws firmly again.

·Place the sieve stack in the middle of the car-

riage and loosen the two clamping screws.

·Push the carriage upwards until the top edge of

the sieve stack is at the same level as the guide
ring. Then tighten the clamping screws again.

·Lift the tapping arm slightly and place the sieve

lid E with cork plug on the top sieve.

C

E

KH