Emissions

Noise characteristics:
Measurement of noise in accordance with DIN 45635-031-01-KL3
The noise characteristics depend on the set amplitude or sieve
base acceleration, the number of sieves and the type of the
material being sieved.

Example 1:
Workplace-related emissions LpAeq = = 79 dB(A)
Operating conditions :
Sieved material = quartz sand, grain size <1mm, 5 sieves

Example 2:
Workplace-related emission level LpAeq = 65 dB(A)
Operating conditions :
Sieved material = quartz sand, grain size <1mm, 5 sieves
With noise insulating hood

Maximum loading
max. sieved quantity of sieved material = 3 kg
max. weight of a sieve stack = 6 kg
max. height of a sieve stack: 6 sieves (50 mm) + 1 collecting base
(50 mm)

Systems of protection
IP 54 and IP 20 resp. in the area of the sieve holder passage

Device dimensions
Width: 715 mm, height: approx. 760 mm, depth: 520 mm
Weight: approx. 92 kg (without sieve stack)

Required standing area
800mm x 550mm; no safety spacing needed.

Technical data
Number of oscillations of the sieve plate: 280/min.