Displaying and setting the alarm points – RKI Instruments GX-2001 Manual User Manual

Page 26

Advertising
background image

21 • Displaying and Setting the Alarm Points Model GX-2001 Operator’s Man-

Displaying and Setting the Alarm
Points

The Model GX-2001 allows you to display and set the alarm points of the
four target gasses. Each of the target gasses has both a low (Alarm 1) and
high (Alarm 2) alarm setting. Two gasses, CO and H

2

S, have STEL and

TWA alarm points. The alarm point for oxygen is measured as a
percentage of oxygen in the atmosphere. The alarm point for HC is
measured as a percentage of the lower explosive limit (% LEL), and CO
and H

2

S alarm points are measured in parts per million (ppm). The alarm

points and their factory settings are summarized in the table below:

In the table above, Alarm 1 and Alarm 2 for HC, CO, and H

2

S refer to a

rising concentration of these gasses. The first alarm is triggered at the
Alarm 1 concentration listed in the table. For HC that would be 10% LEL,
for CO that would be 25 ppm, and for H

2

S that would be 10 ppm. The

second alarm is triggered when the Alarm 2 concentration is reached. For
oxygen, Alarm 1 is triggered when the concentration of oxygen falls below
19.5%. When the concentration of oxygen rises above 23.5%, Alarm 2 is
activated.

Table 5: Alarm Points of the Four Target Gasses

Gas

Alarm 1

Alarm 2

STEL

TWA

O

2

19.5%

Decreasing

23.5%

None

None

HC

Combustible

s

10% LEL

50% LEL

None

None

CO

25 ppm

50 ppm

200 ppm

25 ppm

H

2

S

10 ppm

30 ppm

15 ppm

10 ppm

Advertising