Model GX-2001 Operator’s Manual

Displaying and Setting the Alarm Points • 22

Display and Adjust the Alarm Points

1. Make sure the Model GX-2001 is turned off. The LCD should be blank.

2. Press and hold the AIR button, then press and hold the MODE/POWER

button.

3. As soon as segments appear on the display (approximately one

second), release the AIR button. When the unit “beeps,” release the
MODE/POWER button to put the Model GX-2001 into Alarm Point
Adjustment Mode. The LCD should show the Alarm 1 setting for O

2

and

the battery level.

NOTE: If the LCD should show “CAL,” the Model GX-2001 is in

Calibration Mode. You will need to press and hold the MODE/
POWER button to turn off the unit. Begin again with step 2 above.

4. Use the MODE/POWER button to cycle through the alarm points of the

target gasses. Oxygen is displayed first followed by H

2

S, CO, then HC.

Use the AIR button to increase the alarm point and the DISP button to
decrease the alarm point. When using the MODE/POWER button,
press it once to display the next reading. If you press and hold the
MODE/POWER button, you may turn off the unit.

NOTE: You can only cycle through the target gasses once before the

Model GX-2001 goes into Start-up Mode followed by Measuring
Mode. If you want to cycle through the gasses again, press and
hold the MODE/POWER button to turn off the Model GX-2001.
Then begin with step 2 above to put the unit back into Alarm Point
Adjustment Mode.

5. When you are finished viewing or adjusting the alarm point settings,

press the MODE/POWER button repeatedly until the ROM number for
your unit appears on the LCD. (the ROM is the component that contains
the software that runs the Model GX-2001.) The Model GX-2001 will
then go into Start-up Mode followed by Measuring Mode.

O2

1 9 . 5

L O

%

Target gas

Battery

charge

Gas concentration

Figure 4: LCD in Alarm Adjustment Mode

Alarm point

.