Replacing the scrubber filters – RKI Instruments GX-2001 Manual User Manual

Page 40

Advertising
background image

35 • Maintenance

Model GX-2001 Operator’s Manual

Replacing the Scrubber Filters

1. Verify that the Model GX-2001 is off.

2. Unscrew the two screws that secure the sensor retainer and sensor

cover retainer to the Model GX-2001 and remove them.

3. Using a small flat blade screwdriver, gently pry each side of the sensor

cover retainer away from the sensor retainer.

4. Remove the sensor cover to gain access to the scrubbers.

5. The charcoal filter disk is located in the CO sensor diffusion port. One

filter disk is used. One H

2

S filter disk is also used in the CO sensor

diffusion port. It is installed over the charcoal filter disk.

One H

2

S removal filter disk is also located in the combustible gas

diffusion port.

6. When the appropriate scrubber filters have been replaced, reinstall the

sensor cover.

7. Reattach the sensor cover retainer in its original position. It should snap

into place onto the sensor retainer.

8. Replace the two screws that secure the sensor retainer and sensor

cover retainer to the Model GX-2001, which you removed in step 2
above.

H2S Removal Filter
Disk

RKI

0

2

CO

MODE DISP AIR

H

2

S

POWER

GX-2001

HC

Sensor Cover

H2S Removal Filter Disk

Charcoal Filter Disk

Sensor Cover Retainer

Figure 9: Removing the Sensor Retainer and Replacing the Scrubber Filters