Specifications – RKI Instruments GX-2001 Manual User Manual

Page 7

Advertising
background image

Model GX-2001 Operator’s Manual

Specifications • 2

Specifications

Table 1: Model GX-2001 Specifications

Model Name

Model GX-2001

Target Gas

Combustibles

(Methane

calibration

standard)

Oxygen (O

2

)

Hydrogen

Sulfide

(H

2

S)

Carbon

Monoxide

(CO)

Range

(Increment)

0-100% LEL

(1% LEL)

0-40.0 vol%

(0.1 vol%)

0-100 ppm

(0.5 ppm)

0-500 ppm

(1 ppm)

Detection

Method

Diffusion

Response

Time

T90 Within 30 Seconds

Display

7-Segment Digital LCD

Gas Alarms

(Factory

Settings)

Alarm 1: 10%
LEL
Alarm 2: 50%
LEL
Over: 100%
LEL

Alarm 1: 19.5
vol%
(Decreasing)
Alarm 2: 23.5
vol%
Over: 40.0
vol%

Alarm 1: 10
ppm
Alarm 2: 30
ppm
TWA: 10 ppm
STEL: 15 ppm

Alarm 1: 25
ppm
Alarm 2: 50
ppm
TWA: 25 ppm
STEL: 200
ppm

Operating

Temperature

and Humidity

-10 to 40

o

C/Below 85% RH (Without Condensation)

Safety/

Regulatory

C US

186718

CSA classified, “C/US”, as Intrinsically Safe. Exia. Class I, Groups A,

B, C, &D. Class I, Zone 0, Group IIC. Temperature Code T3C.

Power Supply

Ni-cad or Ni-MH Battery Pack (2.4 VDC Nominal), Direct Charging

Continuous

Operating

Hours

Up to 10 Hours (Non Alarm Operation After Full Charge), Ni-Cad

Up to 18 Hours (Non Alarm Operation After Full Charge), Ni-MH

Case

High-impact Plastic, RF Shielded, Dust and Weather Proof

Required

Accessories

Charger Station or Charger/Downloading Station w/Software &

Cable