Start up – RKI Instruments GX-2001 Canadian Version User Manual

Page 14

Advertising
background image

14 • Start Up

Model GX-2001 Operator’s Manual

Start Up

This section explains how to start up the Model GX-2001 and to get it ready
for operation.

Start-up Procedure

1. Find a fresh-air environment. This is an environment free of toxic or

combustible gasses and of normal oxygen content (20.9%).

WARNING: The GX-2001 performs an auto zero function when it is

turned on. If the GX-2001 is not in a fresh air environment
when turned on and the auto zero function is performed
with target gases present, the gas readings and alarms
will not be accurate or reliable.

2. Press and hold the MODE/POWER button for one second to turn on the

Model GX-2001 and to activate its LCD. The alarm lights flash, the
buzzer sounds, and the vibrator vibrates.

3. The Model GX-2001 tests all elements of the LCD, then displays the

year, month, day, and time before displaying the battery voltage. The
buzzer sounds again after the battery voltage is displayed.

CAUTION: If the battery level is low or if the battery icon is flashing and

the unit’s buzzer sounds, recharge the Model GX-2001’s
battery.

4. The sensors begin functioning and the GX-2001 then performs an auto

zero function and sets the fresh air readings for all active channels
while the battery voltage is being displayed. No indication is given on
the display regarding the auto zero adjustment.

5. The concentrations of the target gases are displayed on the LCD in the

following order: O

2

, H

2

S, CO, and HC. The Model GX-2001 is now in

Measuring Mode. Oxygen is displayed as a percentage (%), H

2

S and

CO as parts per million (ppm), and HC (combustible gases) as a
percentage of the lower explosive limit (% LEL).

NOTE: You can test the oxygen sensor and alarm circuit by exhaling into

the oxygen sensor. When the LCD reading falls below the alarm
point for oxygen, the buzzer will sound, the vibrator will activate,
the LCD reading for oxygen will flash, and the alarm lights will turn
on.

Advertising