RKI Instruments GX-2001 Canadian Version User Manual

Page 21

Advertising
background image

Model GX-2001 Operator’s Manual

Operation • 21

Resetting Alarms

The GX-2001 has self resetting alarms. A gas alarm will reset and all alarm
indications will stop when the gas concentration decreases below the
alarm level (or rises above for the oxygen alarm 1).

Responding to Alarms

This section describes response to gas, over range, battery, and sensor
failure alarms.

Responding to Gas Alarms

1. Determine which gas alarm has been activated.

2. Follow your established procedure for an increasing gas condition or a

decreasing oxygen condition.

Responding to Over Range Alarms

WARNING: An over range condition may indicate an extreme

combustible gas, toxic gas, or oxygen content condition.
Confirm a normal condition with a different Model GX-
2001 or with another gas detecting device.

1. Determine which gas alarm has been activated.

2. Follow your established procedure for an increasing gas condition.

3. Calibrate the Model GX-2001 as described in the calibration section of

this manual.

4. If the over range condition continues, you may need to replace the

sensor that has triggered the over range alarm.

5. If the over range condition continues after you have replaced the

sensor, contact RKI Instruments, Inc. for further instructions.

Responding to Battery Alarms

WARNING: The Model GX-2001 is not operational as a gas

monitoring device during a dead battery alarm. Take the
Model GX-2001 to a non-hazardous area and recharge the
battery as described in “Recharging the Battery Pack.”

The Model GX-2001 is fully functional during a low battery warning.
However, depending upon conditions, the Model GX-2001 has only two-to-
three hours of operating time left after the low level battery alarm has been
triggered. Recharge the battery pack as soon as possible as described in
“Recharging the Battery Pack.”

Advertising