Replacing the scrubber filters – RKI Instruments GX-2001 Canadian Version User Manual

Page 38

Advertising
background image

38 • Maintenance

Model GX-2001 Operator’s Manual

Replacing the Scrubber Filters

1. Verify that the Model GX-2001 is off.

2. Unscrew the two screws that secure the sensor retainer and sensor

cover retainer to the Model GX-2001 and remove them.

3. Using a small flat blade screwdriver, gently pry each side of the sensor

cover retainer away from the sensor retainer.

4. Remove the sensor cover to gain access to the scrubbers.

5. The charcoal filter disk is located in the CO sensor diffusion port. One

filter disk is used. Two H2S filter disks are also used in the CO sensor
diffusion port. They are installed over the charcoal filter disk.

Two H2S removal filter disks are also located in the combustible gas
diffusion port.

6. When the appropriate scrubber filters have been replaced, reinstall the

sensor cover.

7. Reattach the sensor cover retainer in its original position. It should snap

into place onto the sensor retainer.

8. Replace the two screws that secure the sensor retainer and sensor

cover retainer to the Model GX-2001, which you removed in step 2
above.

H2S Removal Filter
Disk

0

2

RKI

POWER

H

2

S

MODE

DISP

AIR

CO

H2S Removal Filter Disk

Sensor Cover

HC

GX-2001

Sensor Cover Retainer

Charcoal Filter Disk

Figure 9: Removing the Sensor Retainer and Replacing the Scrubber Filters

Advertising