Using Maintenance Mode • 119
button to clear the alarm indications.
1. Use the AIRS or (SHIFT)T buttons to display the ALM SLNC menu
item.
2. Press and release the POWER ENTER button. The screen will indicate
the current setting.
3. Use the AIRS or (SHIFT)T button to display the desired setting.
4. Press and release the POWER ENTER button to save the setting and
return to the main menu.
Updating the Data Log Interval Setting
This setting indicates how often the GX-2012 saves readings to the data
logger. The following interval times can be selected: 6 minutes, 5 minutes, 3
minutes, 1 minute, 30 seconds, 20 seconds, or 10 seconds. The factory setting
is 5 minutes (300 seconds).
1. Use the AIRS or (SHIFT)T buttons to display the LG INTVL menu
item.
MAINTENANCE
ALM SLNC
On
MAINTENANCE
ALM SLNC
MAINTENANCE
LG INTVL