176 • Bump Testing

GX-2012 Operator’s Manual

calibration initiated because of a failed bump test is the difference
between the bump test gas time and the calibration time defined in the
BUMP--SET menu item in Maintenance Mode.

•

At the end of the calibration, the instrument displays the results from
both the bump test and the calibration. Use the AIRS or (SHIFT)T
buttons to scroll between the calibration/bump test results, the bump
test gas readings, and the calibration gas readings.

•

To return to the BUMP screen in Calibration Mode, press POWER

1
20.9
37
10.0

%LEL

%

ppm

ppm

MAINTENANCE

CH

4

O

2

CAL 60

CO

H

2

S

FP
PP
FF

PP

%LEL

%

ppm

ppm

MAINTENANCE

CH

4

O

2

BUMP/CAL

CO

H

2

S

1
12.0

5
23.0

%LEL

%

ppm

ppm

MAINTENANCE

CH

4

O

2

BUMP/

CO

H

2

S

49
12.0

5
25.0

%LEL

%

ppm

ppm

MAINTENANCE

CH

4

O

2

/CAL

CO

H

2

S