Portugal – RKI Instruments GX-2012 Manual User Manual
Page 147

GX-2012 Operator’s Manual
Using Maintenance Mode • 139
1. Use the AIRS or (SHIFT)T buttons to display the PORTUGAL menu
item.
2. Press and release the POWER ENTER button. The screen will indicate
the current setting.
3. Use the AIRS or (SHIFT)T button to display the desired setting.
4. Press and release the POWER ENTER button to save the setting and
return to the main menu.
Updating the Leak Check Mode CO
Channel Display Setting
With CO DISP set to OFF (factory setting), the CO channel is not displayed
in Leak Check Mode but is still displayed in Normal Mode.
With CO DISP set to On, the CO channel will be displayed in Leak Check
Mode but there will be no CO alarms and the CO channel reading will not be
saved to any snap log data taken during the operating session.
MAINTENANCE
PORTUGAL
OFF
MAINTENANCE
PORTUGAL