48 • Display Mode
GX-2012 Operator’s Manual
Date/Time Screen
The date/time screen displays the current date and time with the order being
year, month, day, time.
Remaining Log Time Screen
The remaining log time screen displays the time remaining until the data
logger memory is full. The remaining log time depends on how often the GX-
2012 stores data to the data log, how many channels are active, and how often
the GX-2012 is turned on and off.
The table below illustrates how much data logging time is available for the
various interval times. It assumes that the unit is set up with all 5 sensors and
there are no alarm occurrences.
Table 6: Data Logging Capacity, 5 Sensor GX-2012
Interval
Time
Data Logging Time
10 seconds
10 hours
20 seconds
20 hours
30 seconds
30 hours
1 minute
60 hours
3 minutes
180 hours
5 minutes
300 hours
10 minutes
600 hours