O n 30 30 – RKI Instruments SDM-2009 User Manual

Page 22

Advertising
background image

18 • Setting the Operational Parameters in Edit Mode

4.

If you wish to cancel setting the bump test parameters or were just viewing the
parameters to confirm their values, press and release the BUMP

T

button to

return to the home screen.

To continue and change parameters, press and release the EDIT ENTER
button. The Air Sample Time parameter will be displayed. For a description of
the Air Sample Time bump test parameter, see “Bump Test & Calibration
Parameters” on page 14.

5.

Use the BUMP

T

and CAL

S

buttons to scroll through the different bump test

parameters. Do not scroll past the AUTO--C parameter until you have finished
making any desired changes.

6.

To edit a parameter, press and release the EDIT ENTER button when that
parameter is displayed. An “E” will appear to the left of the parameter name
indicating that the parameter value can be changed. In the example below, the
AIR--IN parameter has been selected for updating.

7.

Use the BUMP

T

and CAL

S

buttons to set the parameter to the desired

value, then press and release the EDIT ENTER button. The “E” to the right of
parameter name will disappear.

8.

Repeat step 6 - step 7 to set any other parameters.

30

AIR--IN

30

AIR--IN

30

GAS--IN

CHECK

50

AUTO--C

O n

30 30

DECISION

50 O n

50 12.0

TRANSMIT

CO ppm H2S ppm

CH4 %LEL OXY vol%

50 25.0

30

E AIR--IN