Air gas, Decision, Transmit – RKI Instruments SDM-2009 User Manual

Page 23: Bump/cal

Advertising
background image

Setting the Operational Parameters in Edit Mode • 19

9.

When you are done setting the parameters, use the BUMP

T

button to scroll

past the AUTO--C parameter screen. The screen will indicate that the
parameter changes have been saved.

The instrument will return to the home screen.

Setting the Calibration Parameters

Do the following to set the calibration parameters.

1.

Turn on the SDM-2009 with an instrument and establish a connection between
them as described above in "Turning on the SDM-2009 with an Instrument".
The instrument will display the home screen.

2.

Press and hold the EDIT ENTER button for 2 seconds. The instrument will
display the Edit Mode screen.

3.

Press and release the CAL

S

button. The instrument will alternate between

the following screens that display the two calibration parameters and their
settings.

4.

If you wish to cancel setting the calibration parameters or were just viewing the
parameters to confirm their values, press and release the CAL

S

button to

return to the home screen.

To continue and change parameters, press and release the EDIT ENTER
button. The Air Sample Time parameter will be displayed. For a description of
the Air Sample Time bump test parameter, see “Bump Test & Calibration

30 30

DECISION

50 O n

50 12.0

TRANSMIT

CO ppm H2S ppm

CH4 %LEL OXY vol%

50 25.0

BUMP/CAL

30 90

CAL

Air GAS

CAL

Advertising