Troubleshooting • 35

• No IR connection

between
instrument and
SDM-2009

• SDM-2009 is not turned

on.

• The instrument is not

correctly inserted into the
instrument cradle.

• The IR window is dirty.

1. Turn on the SDM-2009. If it does

not turn on, check that the AC
adapter is plugged into an AC
socket and to the jack on the
back of the SDM-2009.

2. Check to make sure the

instrument is inserted properly.

3. Clean the IR window on the

SDM-2009 and the GX-2009.

4. If the difficulties continue, contact

RKI Instruments, Inc. for further
instruction.

• Control buttons do

not function
correctly for
standalone
operation

• The SDM-2009 is

connected to a computer
and the PC Controller
program is running.

Disconnect the USB cable from the
back of the SDM-2009 if it is
connected. See the SDM-2009 PC
Controlled Configuration Operator’s
Manual for instructions to use the
SDM-2009 in PC Controlled
configuration.

• The COPY

function does not
copy files to the
USB drive

• The USB drive is

defective.

• The USB drive is not

inserted properly.

• The SDM-2009 is

connected to a computer,
the PC Controller program
is running, and an
instrument is turned on.

1. Perform a COPY operation with a

different USB drive.

2. Be sure that the USB drive is

inserted correctly.

3. Disconnect the SDM-2009 from

the computer.

4. If the difficulties continue, contact

RKI Instruments, Inc. for further
instruction.

• The rechargeable

batteries in the
instrument are not
charging

• The instrument is not

properly inserted in the
instrument cradle.

• The charger is

malfunctioning.

• The batteries are faulty.

1. Be sure that the instrument is

fully inserted in the instrument
cradle.

2. Try to charge the batteries in a

BC-2009 charger if one is
available to determine if the
SDM-2009 is at fault.

3. Try to charge the batteries using

a different SDM-2009 if one is
available to determine if the
SDM-2009 is at fault.

4. Replace the batteries and

attempt to charge them. See the
GX-2009 Operator’s Manual for
instructions to change the
batteries.

5. If the difficulties continue, contact

RKI Instruments, Inc. for further
instruction.