16 • Overview

Chapter 3: Hardware Setup and Software Installation

Overview

There are two parts to preparing the SM-2001U and SM-2003U for use:
hardware setup and software installation. This chapter describes how to
setup the SM-2001U and SM-2003U hardware. It also describes how to
install the Single Module Data Viewer Software on a Windows based
personal computer.

Hardware Setup

The hardware setup consists of some parts assembly, connecting
calibration gas, and setting the MODE switch to the position that
corresponds the desired bump test parameter settings.

Parts Assembly

Some assembly is required before the SM-2001U or SM-2003U can be
used. Perform the following assembly before attempting to use a
calibration station.

1. If you have an SM-2001U, install the instrument sample adapter.

Push the coiled end of the tubing onto the onto the recessed stub
located at the rear top of the calibration station and the straight end
of the tubing onto the recessed stub located on the upper left front
of the calibration station.

If you have an SM-2003U, install the green tube with the plastic
push-on fitting. Push the open end of the tube onto the recessed
stub located on the top rear of the calibration station.

2. Install the air filter. Push the open end of the tube on the filter onto

the AIR fitting on the back of the calibration station.

3. If it is not already installed, install the exhaust plug on the bottom

exhaust port. It is a black rubber plug that fits over the port.

4. Plug the power cord into the power cord connector on the back of

the calibration station.