34 • Calibration and Bump Test Records

Copying Calibration and Bump Test Records to a USB
Flash Drive

The calibration station will only perform a copy operation if there is at
least one record saved in it’s memory. If there are no records saved in the
station’s memory, the COPY LED will be off and the COPY button will not
function. Do the following to save calibration and bump test records in the
calibration station’s memory to a USB flash drive.

1. If the calibration station is off, turn it on using the power switch on

the back of the station.

NOTE: The USB port on the front of the calibration station

cannot be used to connect the SM-2001U or the SM-
2003U to a computer, only to save calibration and bump
test records to a USB flash drive.

2. Install a USB flash drive into the USB port on the front of the

calibration station. The calibration station will take a few seconds to
determine how much memory is available in the flash drive.

If the flash drive’s available memory it is not enough for the
contents of the calibration station’s memory, the COPY LED will
alternate between green and red. Enough memory will have to
be cleared in the flash drive for the records in the calibration
station’s memory.

If there is enough available memory in the flash drive for the
contents of the calibration station’s memory, the COPY LED will
begin flashing in the same color that it was before installing the
flash drive.

3. Press and release the COPY button. The COPY LED will become

solid red while the records in the calibration station’s memory are
copied to the flash drive.

Advertising