Calibration and Bump Test Records • 33

Calibration and Bump Test Records

The SM-2001U and SM-2003U save a record of each bump test and
calibration performed. The calibration stations are capable of saving up
to 200 such records. When a calibration station’s memory becomes full,
the oldest record is overwritten when a new record is saved. These
saved records can be saved to a USB flash drive using the USB port.

Available Memory in the Calibration Station

The COPY function LED above the copy button indicates how much of
the calibration station’s memory has been used. The table below
describes the various indications.

The calibration station’s memory can be cleared by simultaneously
pressing and holding the CAL. and COPY control buttons for five
seconds.

Table 5: COPY LED Indications

COPY LED Indication

Memory Used

Off

None. No records are saved

Solid Green

Less than 80% of the calibration sta-
tion’s memory has been used.

Solid Yellow

More than 80% of the calibration sta-
tion’s memory has been used.

Solid Red

The calibration station’s memory is
full. A newly saved record will over-
write the oldest one.