20 • Hardware Setup
The parameter settings for each MODE switch position are shown in
Table 4 below.
Set the MODE switch to the position that corresponds to the desired
parameter settings using a small flat blade screwdriver.
Table 4: MODE Switch Parameter Settings
Switch
Position
Air Purge
Time
Gas Exposure
Time
Tolerance
Auto. Cal.
0
30
seconds
30 seconds
± 50%
Off
1
45 Seconds
± 20%
2
± 30%
3
± 50%
4
90 seconds
± 10%
5
30 seconds
± 50%
On
6
45 seconds
± 20%
7
± 30%
8
90 seconds
± 10%
9
60 seconds
± 20%