20 • Alarm Indications

Beacon 100 Gas Monitor Operator’s Manual

Responding to an alarm 3 condition

This section suggests response to an alarm 3 condition.

1.

Follow your established procedure for a high level combustible or toxic gas condition
or an increasing oxygen content condition.

2.

After the gas reading falls below the alarm 3 setpoint, press the reset switch to reset
the alarm 3 circuit. Resetting the alarm 3 circuit turns off the ALARM 3 light, and
de-energizes the alarm 3 relay.

NOTE:

The alarm 3 condition cannot be silenced. The reset switch must be pressed after
the gas reading falls below the alarm 3 level in order to de-energize the alarm 3
relay.

Fail Condition

This section describes the audible and visual indications for a fail condition and suggests
response to a fail condition.

Fail condition indications

The Beacon 100 senses a fail condition for any of the following:

the detector wiring is disconnected or incorrectly connected

the display reading is below the fail setpoint

the monitor or detector is malfunctioning

When the Beacon 100 senses a fail condition, it alerts you as follows:

the FAIL light turns on

the gas reading flashes

the buzzer sounds a steady tone

the fail relay de-energizes

Responding to a fail condition

This section suggests response to a fail condition.

1.

Verify that the detector wiring is correctly and securely connected.

2.

See the troubleshooting guide in the applicable detector head operator’s manual.