RKI Instruments 35-3001-04 User Manual

Page 10

Advertising
background image

6 • 35-3001-04 Carbon Monoxide Sample-Draw Detector

Figure 3: CO Amplifier Component Location

Zero Pot

The zero pot is located in the lower right corner of the amplifier (see Figure 3). Use a small flat
blade screwdriver to turn the zero pot’s adjustment screw and adjust the amplifier’s zero output
during the start up and calibration procedures. Turn the adjustment screw clockwise to increase the
zero output and counterclockwise to decrease the zero output.

Span Pot

The span pot is located to the left of the zero pot (see Figure 3). Use a small flat blade screwdriver
to turn the span pot’s adjustment screw and adjust the amplifier’s gas response during the
calibration procedure. Turn the adjustment screw clockwise to increase the span output and
counterclockwise to decrease the span output.

Test Points

The test points are on the top edge of the amplifier (see Figure 3). The test points produce a 100 mV
to 500 mV output that corresponds to the sample-draw detector’s 4 to 20 mA output. Use the test
points and a voltmeter to measure the amplifier’s output during the start-up and calibration
procedures. The black test point (CAL 2) in the upper left corner is the negative (-) test point and
the red test point (CAL 1) in the upper right corner is the positive (+) test point.

Zero Pot

Span Pot

Test Points

100-500 mV Range