RKI Instruments 35-3001-04 User Manual

Page 19

Advertising
background image

35-3001-04 Carbon Monoxide Sample-Draw Detector • 15

5.

If the fail condition continues, replace the sensor as described in “Replacing Components of
the Sample-Draw Detector” on page 15.

6.

If the fail condition continues, contact RKI Instruments, Inc. for further instruction.

Slow or No Response/Difficult or Unable to Calibrate

Symptoms

The sensor responds slowly or does not respond during the monthly response test.

Unable to accurately set the zero or response reading during the calibration procedure.

The sensor requires frequent calibration.

Probable causes

The calibration cylinder is low, out-dated, or defective.

If a demand flow regulator calibration kit is used, the demand flow regulator is not functioning
properly.

The sample-draw detector’s flow rate is too low because of an obstructed sample line, failed
pump, etc.

The sample-draw detector is malfunctioning.

Recommended action

1.

Verify that the calibration cylinder contains an adequate supply of a fresh test sample.

2.

If a demand flow regulator calibration kit is used, use a different demand flow regulator to
determine if the original one is functioning properly.

3.

If necessary, set the correct flow rate with the sensor flow control valve.

4.

If you cannot set the correct flow rate, check the sample line for obstructions or kinks.

5.

If the calibration/response difficulties continue, replace the sensor as described later in this
section.

6.

If the calibration/response difficulties continue, contact RKI Instruments, Inc. for further
instruction.

Replacing Components of the Sample-Draw Detector

This section includes procedures to replace the carbon monoxide sensor, CO amplifier,
hydrophobic filter, and particle filter.

Replacing the CO Sensor

1.

Turn off the controller.

2.

Turn off power to the controller.

3.

Open the housing door of the sample-draw detector.

4.

Unscrew the two screws in the upper right and lower left corners of the CO flow block then lift
the CO amplifier and sensor off of the flow block.

There is a gasket at the bottom of the flow block. Be sure the gasket stays in place.

5.

Unplug the sensor from the CO amplifier.

6.

Verify that you are using the correct replacement sensor, then plug the sensor into the CO
amplifier.

7.

Place the sensor and amplifier in the CO sensor cavity.

8.

Secure the CO amplifier on the flow block with the two screws you removed in step 4.

9.

Close and latch the housing door.

Advertising